Projet de comité
ISO/CD 24085-1
Implants chirurgicaux --Prothèses partielles et totales de l'épaule — Partie 1: Détermination de la résistance à torque des têtes humérales
Numéro de référence
ISO/CD 24085-1
Edition 1
Projet de comité
ISO/CD 24085-1
93832
Examen du projet par le comité en cours.

Résumé

This document describes a method of determining the torque required, under specified laboratory conditions, to loosen the fixation of the humeral head of modular anatomic shoulder joint replacement in which the humeral head is not intended to rotate relative to the humeral component. This test method is used to evaluate humeral components of total or partial shoulder joint replacements in which the humeral head and humeral component are secured together by a connector(s) (see Figure A.1) assembled by a compressive load and in which the humeral head and humeral component are separate components, and which are made of metallic or non-metallic materials. There can be one, two or three connectors as part of the modular anatomic shoulder joint replacement. (see Figure A.3). This document does not cover methods of examining the test specimens after testing.

Informations générales

  •  : Projet
    : Projet de comité (CD) enregistré [30.00]
  •  : 1
  • ISO/TC 150/SC 4
  • RSS mises à jour

Vous avez une question?