Diabetes patients will benefit from the latest version of an ISOstandard for self-testing glucose monitoring systems. The standard requiresthat manufacturers of these systems provide even greater accuracy.

The new ISO 15197:2013, In vitro diagnostic test systems - Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus, differs from the previous 2003 version on the following points:

  • Increased accuracy for glucose metersystems, in particular for glucose values greater than 75 mg/dl (4,2 mmol/l)
  • Manufacturers of glucose meter systems mustensure their technology enables accuracy to improve from +-20% to +-15%
  • The new version accounts for 99% ofresults, as opposed to 95% for the previous one
  • For the first time, the standard providesformal acceptance criteria for accuracy as regards testing by patients andassessment of interferents (including hematocrit).

Project leader for ISO 15197, Dr. AlanCariski, comments, "More accurate glucose measurements will help patients tobetter regulate their diabetes through more informed treatment decisions thatmay affect, for example, dietary intake and medication dose, especially insulin."