ISO 14117:2012 Preview

Active implantable medical devices -- Electromagnetic compatibility -- EMC test protocols for implantable cardiac pacemakers, implantable cardioverter defibrillators and cardiac resynchronization devices

ISO 14117:2012 specifies test methodologies for the evaluation of the electromagnetic compatibility (EMC) of active implantable cardiovascular devices that provide one or more therapies for bradycardia, tachycardia and cardiac resynchronization.

It specifies performance limits of these devices, which are subject to interactions with EM emitters operating across the EM spectrum in the two following ranges:

0 Hz ≤ f < 450 MHz;

450 MHz ≤ f ≤ 3 000 MHz.

ISO 14117:2012 also specifies requirements for the protection of these devices from EM fields encountered in a therapeutic environment and defines their required accompanying documentation, providing manufacturers of EM emitters with information about their expected level of immunity.


General information

  • Current status : Published
    Publication date : 2012-07
  • Edition : 1
    Number of pages : 97
  • :
    ISO/TC 150/SC 6
    Active implants
  • 33.100.01
    Electromagnetic compatibility in general
    11.040.40
    Implants for surgery, prosthetics and orthotics