4   Properties of measuring devices

4.1 (3.2)
indication
quantity value provided by a measuring instrument or a measuring system

NOTE 1   An indication may be presented in visual or acoustic form or may be transferred to another device. An indication is often given by the position of a pointer on the display for analog outputs, a displayed or printed number for digital outputs, a code pattern for code outputs, or an assigned quantity value for material measures.

NOTE 2   An indication and a corresponding value of the quantity being measured are not necessarily values of quantities of the same kind.

4.2
blank indication
background indication
indication obtained from a phenomenon, body, or substance similar to the one under investigation, but for which a quantity of interest is supposed not to be present, or is not contributing to the indication
4.3 (4.19)
indication interval
set of quantity values bounded by extreme possible indications

NOTE 1   An indication interval is usually stated in terms of its smallest and greatest quantity values, for example “99 V to 201 V”.

NOTE 2   In some fields, the term is “range of indications”.

4.4 (5.1)
nominal indication interval
nominal interval
set of quantity values, bounded by rounded or approximate extreme indications, obtainable with a particular setting of the controls of a measuring instrument or measuring system and used to designate that setting

NOTE 1   A nominal indication interval is usually stated as its smallest and greatest quantity values, for example “100 V to 200 V”.

NOTE 2   In some fields, the term is “nominal range”.

4.5 (5.2)
range of a nominal indication interval
absolute value of the difference between the extreme quantity values of a nominal indication interval

EXAMPLE   For a nominal indication interval of −10 V to +10 V, the range of the nominal indication interval is 20 V.

NOTE   Range of a nominal indication interval is sometimes termed “span of a nominal interval”.

4.6 (5.3)
nominal quantity value
nominal value
rounded or approximate value of a characterizing quantity of a measuring instrument or measuring system that provides guidance for its appropriate use

EXAMPLE 1   100 Ω as the nominal quantity value marked on a standard resistor.

EXAMPLE 2   1 000 ml as the nominal quantity value marked on a single‑mark volumetric flask.

EXAMPLE 3   0.1 mol/l as the nominal quantity value for amount‑of‑substance concentration of a solution of hydrogen chloride, HCl.

EXAMPLE 4   −20 °C as a maximum Celsius temperature for storage.

NOTE   “Nominal quantity value” and “nominal value” are not to be confused with “nominal property value” (see 1.30, Note 2).

4.7 (5.4)
measuring interval
working interval
set of values of quantities of the same kind that can be measured by a given measuring instrument or measuring system with specified instrumental uncertainty, under defined conditions

NOTE 1   In some fields, the term is “measuring range” or “measurement range”.

NOTE 2   The lower limit of a measuring interval should not be confused with detection limit.

4.8
steady‑state operating condition
operating condition of a measuring instrument or measuring system in which the relation established by calibration remains valid even for a measurand varying with time
4.9 (5.5)
rated operating condition
operating condition that must be fulfilled during measurement in order that a measuring instrument or measuring system perform as designed

NOTE   Rated operating conditions generally specify intervals of values for a quantity being measured and for any influence quantity.

4.10 (5.6)
limiting operating condition
extreme operating condition that a measuring instrument or measuring system is required to withstand without damage, and without degradation of specified metrological properties, when it is subsequently operated under its rated operating conditions

NOTE 1   Limiting conditions for storage, transport or operation can differ.

NOTE 2   Limiting conditions can include limiting values of a quantity being measured and of any influence quantity.

4.11 (5.7)
reference operating condition
reference condition
operating condition prescribed for evaluating the performance of a measuring instrument or measuring system or for comparison of measurement results

NOTE 1   Reference operating conditions specify intervals of values of the measurand and of the influence quantities.

NOTE 2   In IEC 60050‑300, item 311‑06‑02, the term “reference condition” refers to an operating condition under which the specified instrumental measurement uncertainty is the smallest possible.

4.12 (5.10)
sensitivity of a measuring system
sensitivity
quotient of the change in an indication of a measuring system and the corresponding change in a value of a quantity being measured

NOTE 1   Sensitivity of a measuring system can depend on the value of the quantity being measured.

NOTE 2   The change considered in a value of a quantity being measured must be large compared with the resolution.

4.13
selectivity of a measuring system
selectivity
property of a measuring system, used with a specified measurement procedure, whereby it provides measured quantity values for one or more measurands such that the values of each measurand are independent of other measurands or other quantities in the phenomenon, body, or substance being investigated

EXAMPLE 1   Capability of a measuring system including a mass spectrometer to measure the ion current ratio generated by two specified compounds without disturbance by other specified sources of electric current.

EXAMPLE 2   Capability of a measuring system to measure the power of a signal component at a given frequency without being disturbed by signal components or other signals at other frequencies.

EXAMPLE 3   Capability of a receiver to discriminate between a wanted signal and unwanted signals, often having frequencies slightly different from the frequency of the wanted signal.

EXAMPLE 4   Capability of a measuring system for ionizing radiation to respond to a given radiation to be measured in the presence of concomitant radiation.

EXAMPLE 5   Capability of a measuring system to measure the amount‑of‑substance concentration of creatininium in blood plasma by a JaffĂ© procedure without being influenced by the glucose, urate, ketone, and protein concentrations.

EXAMPLE 6   Capability of a mass spectrometer to measure the amount‑of‑substance abundance of the 28Si isotope and of the 30Si isotope in silicon from a geological deposit without influence between the two, or from the 29Si isotope.

NOTE 1   In physics, there is only one measurand; the other quantities are of the same kind as the measurand, and they are input quantities to the measuring system.

NOTE 2   In chemistry, the measured quantities often involve different components in the system undergoing measurement and these quantities are not necessarily of the same kind.

NOTE 3   In chemistry, selectivity of a measuring system is usually obtained for quantities with selected components in concentrations within stated intervals.

NOTE 4   Selectivity as used in physics (see Note 1) is a concept close to specificity as it is sometimes used in chemistry.

4.14
resolution
smallest change in a quantity being measured that causes a perceptible change in the corresponding indication

NOTE   Resolution can depend on, for example, noise (internal or external) or friction. It may also depend on the value of a quantity being measured.

4.15 (5.12)
resolution of a displaying device
smallest difference between displayed indications that can be meaningfully distinguished
4.16 (5.11)
discrimination threshold
largest change in a value of a quantity being measured that causes no detectable change in the corresponding indication

NOTE   Discrimination threshold may depend on, e.g. noise (internal or external) or friction. It can also depend on the value of the quantity being measured and how the change is applied.

4.17 (5.13)
dead band
maximum interval through which a value of a quantity being measured can be changed in both directions without producing a detectable change in the corresponding indication

NOTE   Dead band can depend on the rate of change.

4.18
detection limit
limit of detection
measured quantity value, obtained by a given measurement procedure, for which the probability of falsely claiming the absence of a component in a material is β, given a probability α of falsely claiming its presence

NOTE 1   IUPAC recommends default values for α and β equal to 0.05.

NOTE 2   The abbreviation LOD is sometimes used.

NOTE 3   The term “sensitivity” is discouraged for ‘detection limit’.

4.19 (5.14)
stability of a measuring instrument
stability
property of a measuring instrument, whereby its metrological properties remain constant in time

NOTE   Stability may be quantified in several ways.

EXAMPLE 1   In terms of the duration of a time interval over which a metrological property changes by a stated amount.

EXAMPLE 2   In terms of the change of a property over a stated time interval.

4.20 (5.25)
instrumental bias
average of replicate indications minus a reference quantity value
4.21 (5.16)
instrumental drift
continuous or incremental change over time in indication, due to changes in metrological properties of a measuring instrument

NOTE   Instrumental drift is related neither to a change in a quantity being measured nor to a change of any recognized influence quantity.

4.22
variation due to an influence quantity
difference in indication for a given measured quantity value, or in quantity values supplied by a material measure, when an influence quantity assumes successively two different quantity values
4.23 (5.17)
step response time
duration between the instant when an input quantity value of a measuring instrument or measuring system is subjected to an abrupt change between two specified constant quantity values and the instant when a corresponding indication settles within specified limits around its final steady value
4.24
instrumental measurement uncertainty
component of measurement uncertainty arising from a measuring instrument or measuring system in use

NOTE 1   Instrumental measurement uncertainty is obtained through calibration of a measuring instrument or measuring system, except for a primary measurement standard for which other means are used.

NOTE 2   Instrumental uncertainty is used in a Type B evaluation of measurement uncertainty.

NOTE 3   Information relevant to instrumental measurement uncertainty may be given in the instrument specifications.

4.25 (5.19)
accuracy class
class of measuring instruments or measuring systems that meet stated metrological requirements that are intended to keep measurement errors or instrumental uncertainties within specified limits under specified operating conditions

NOTE 1   An accuracy class is usually denoted by a number or symbol adopted by convention.

NOTE 2   Accuracy class applies to material measures.

4.26 (5.21)
maximum permissible measurement error
maximum permissible error
limit of error
extreme value of measurement error, with respect to a known reference quantity value, permitted by specifications or regulations for a given measurement, measuring instrument, or measuring system

NOTE 1   Usually, the term “maximum permissible errors” or “limits of error” is used where there are two extreme values.

NOTE 2   The term “tolerance” should not be used to designate ‘maximum permissible error’.

4.27 (5.22)
datum measurement error
datum error
measurement error of a measuring instrument or measuring system at a specified measured quantity value
4.28 (5.23)
zero error
datum measurement error where the specified measured quantity value is zero

NOTE   Zero error should not be confused with absence of measurement error.

4.29
null measurement uncertainty
measurement uncertainty where the specified measured quantity value is zero

NOTE 1   Null measurement uncertainty is associated with a null or near zero indication and covers an interval where one does not know whether the measurand is too small to be detected or the indication of the measuring instrument is due only to noise.

NOTE 2   The concept of ‘null measurement uncertainty’ also applies when a difference is obtained between measurement of a sample and a blank.

4.30
calibration diagram
graphical expression of the relation between indication and corresponding measurement result

NOTE 1   A calibration diagram is the strip of the plane defined by the axis of the indication and the axis of measurement result, that represents the relation between an indication and a set of measured quantity values. A one‑to‑many relation is given, and the width of the strip for a given indication provides the instrumental measurement uncertainty.

NOTE 2   Alternative expressions of the relation include a calibration curve and associated measurement uncertainty, a calibration table, or a set of functions.

NOTE 3   This concept pertains to a calibration when the instrumental measurement uncertainty is large in comparison with the measurement uncertainties associated with the quantity values of measurement standards.

4.31
calibration curve
expression of the relation between indication and corresponding measured quantity value

NOTE   A calibration curve expresses a one‑to‑one relation that does not supply a measurement result as it bears no information about the measurement uncertainty.