International Standard
ISO 8932-2:2026
Meteorology — Radiosonde — Part 2: Laboratory test method for errors in radiosonde humidity sensor calibration
Reference number
ISO 8932-2:2026
Edition 1
2026-05
Preview
ISO 8932-2:2026
84235
No disponible en español
Publicado (Edición 1, 2026)

ISO 8932-2:2026

ISO 8932-2:2026
84235
Idioma
Formato
CHF 135

Resumen

The document specifies testing procedures for determining calibration error for radiosonde humidity sensors sampled from mass production batches based on varying the levels of relative humidity at atmospheric upper-air temperatures using a laboratory setup. This document provides:

  1. technical requirements for a laboratory setup to evaluate the calibration errors of radiosonde humidity measurement;
  2. a test procedure for evaluating calibration error of radiosonde humidity sensors for a temperature range1) of −90 °C to 35 °C and for a relative humidity of 1 %rh to 100 %rh. Note, this document, is based upon relative humidity calculated by the percentage of water vapour pressure divided by saturation water vapour pressure over liquid water, not over ice, even at temperatures below 0 °C; hence, the maximum relative humidity is less than 100 %rh below 0 °C;
  3. a method for evaluating the uncertainty for the measured radiosonde humidity calibration errors.

1) Currently, the lowest possible temperature of commercially-available climate chambers is approximately -75 °C. The temperature range can be adjusted based on the capability of the climate chamber used.